Improvement in Proximal Aortic Endograft Fixation: An Experimental Study Using Different Stent-Grafts in Human Cadaveric Aortas

2003 
Purpose:To compare the proximal fixation characteristics of different types of stent-grafts (SG) and the adjunctive effect of a proximally placed Palmaz stent.Methods:Human cadaveric aortas were obtained at autopsy and cut into 30 aortic segments. Appropriately oversized (10% to 20%) commercial stent-grafts (Zenith, Ancure, Excluder, Talent) and 4 configurations of homemade Palmaz-based stent-grafts (polytetrafluoroethylene or polyester coverings each with no uncovered proximal edge or a 20-mm bare section of stent) were implanted 20 mm into an aortic segment and balloon dilated. Each segment was placed in an experimental apparatus to measure the dislodgment force. Bare Palmaz stents were used to reinforce the proximal fixation of the commercial stent-grafts, and the measurements were repeated. Manual anastomoses were made and their dislodgment force tested for comparison.Results:The median dislodgment force ranged from 6.5 N for the Excluder to 26.5 N for the Zenith (8.0 N for the Talent, 11.8 N for the ...