Piezoelectric Quartz Crystal Microbalance Sensor for Trace Aqueous Cyanide Ion Determination
2007
Using selective reaction chemistry, our present research has developed an online, real-time sensor capable of monitoring toxic cyanide at both drinking water standard and environmental regulatory concentrations. Through the use of a flow cell, aqueous samples containing cyanide are reacted with a gold electrode of a piezoelectric crystal to indirectly sense cyanide concentration by the dissolution of metallic gold. The quartz crystal is an AT-cut wafer sandwiched between two neoprene O-rings within the liquid flow cell. The presence of cyanide in solution results in the selective formation of a soluble dicyano−gold complex according to the Elsner reaction: 4Au + 8CN- + 2H2O + O2 ⇄ 4Au(CN)2- + 4OH-. The resulting loss of gold from the electrode is detected by the piezoelectric crystal as a resonant frequency change. Since free cyanide is a weak acid (pKa = 9.3), available protons compete for cyanide ligands. Therefore, increased sample pH provides higher sensitivity. The detection limits at pH 12 are 16.1...
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
15
References
63
Citations
NaN
KQI