Determination of 226Ra in Urine Using Triple Quadrupole Inductively Coupled Plasma Mass Spectrometry.

2020 
Measuring 226Ra in urine at low levels is critical for both biomonitoring and radiological emergency response. Here we report a new analytical method to quantify 226Ra, as developed and validated by a simple dilute-and-shoot procedure, followed by Inductively Coupled Plasma-triple quadrupole-mass spectrometry detection using 'No Gas MS-MS' mode. The method provides rapid and accurate results for 226Ra with a limit of detection (LOD) down to 0.007 ng/l (0.26 Bq/l). This LOD is well below the recommended action levels for 226Ra detection in children and pregnant women (C/P) set by the Clinical Decision Guide (NCRP Report #161). Results for 226Ra obtained by this method are within ±7.0% of the target values of standard reference materials spiked in the urine.