ITS-90 Scale Realization on the New Radiation Thermometer Calibration Facility at NMi VSL

2008 
In the first half of 2005, Nederlands Meetinstituut Van Swinden Laboratorium B.V. (NMi VSL) redesigned their facilities for radiation thermometry in a new laboratory building and an opportunity arose to implement new measurement methods. The new facility is used for ITS-90 realization and dissemination in the temperature range from − 50 °C to 3,000 °C. A study was performed to compare a silver-point realization with a fixed-point blackbody radiator (FP-BBR) to a sodium heat-pipe blackbody radiator (HP-BBR) traceable via a HTSPRT to a contact thermometry silver point. It was found that the fixed-point realization transfer to the sodium heat pipe results in an uncertainty from 0.2 K to 2.4 K for the ITS-90 over the temperature range from 961.78 °C to 3,000 °C.