Current challenges in personal dosimetry at the US DOE Hanford site.

2002 
An overview is presented of the dosimetry system, dose equivalent calculation methodology, and QA/QC practices used at the US Department of Energy Hanford site. It describes some of the problems encountered in accurately measuring dose equivalent quantities under a broad range of field conditions that do not necessarily correlate with laboratory calibration conditions and the approach taken to solve these problems. Personnel at Hanford are monitored with a combination of Harshaw model 8825 and 8816 thermoluminescence dosemeters and CR-39 etched track dosemeters. Extremities are monitored using the ICN MeasuRing loaded with a Harshaw XD740 chipstrate TLD. All dosemeters employ LiF:Mg,Ti elements that are read on-site with Harshaw model 8800 and 6600 TLD readers. CR-39 dosemeters are electrochemically etched in non-commercial etch chambers and counted with an automated track counting system developed by Pacific Northwest National Laboratory. Problems with over response of the 8825 with respect to H p (0.07), under-response of the 8825 with respect to H p (3), and over response of the 8825 with respect to H p (10) in Hanford's 90 Sr/ 90 Y beta radiation fields are discussed. Approaches to measurement of the operational quantities for field conditions and algorithm solutions to the above problems are described. Methods used to calibrate the ring dosemeter for Hanford field conditions together with limitations of the ring dosemeter in measuring H p (0.07) for extremities, particularly when covered with protective clothing, are also discussed.