Japanese Survey of Radiation Dose Associated With Coronary Computed Tomography Angiography ― 2013 Data From a Multicenter Registry in Daily Practice ―

2020 
BACKGROUND: Although coronary computed tomography angiography (CTA) is frequently used for identifying coronary artery disease, no studies have investigated the radiation dose in detail in Japan. The aim of this study was to estimate the radiation dose of coronary CTA in Japanese clinical practice and to identify the independent predictors associated with radiation dose.Methods and Results:A multicenter, retrospective, observational study (54 institutions) was conducted for estimating the radiation dose of coronary CTA in 2,469 patients between January and December 2013. Independent predictors associated with radiation dose were investigated on linear regression analysis. Median dose-length product (DLP) was 809.0 mGy.cm (IQR, 350.0-1,368.8 mGy.cm), corresponding to an estimated radiation dose of 11 mSv. The DLP per site significantly differed between institutions (median DLP per site, 92-2,131 mGy.cm; P<0.05). Independent predictors associated with radiation dose on multivariable linear regression were body weight, heart rate, non-stable sinus rhythm, scan length, tube voltage setting, electrocardiogram (ECG)-gated scanning protocol, and the image reconstruction technique (P<0.05 each). CONCLUSIONS: The coronary CTA radiation dose was relatively high in 2013, and it varied significantly between institutions. Effective strategies for radiation dose reduction were low tube voltage