Morphological and functional consequences of transradial coronary angiography on the radial artery: implications for its use as a bypass conduit †

2014 
OBJECTIVES: To assess the degree of damage to the radial artery (RA) in coronary artery bypass grafting (CABG) patients who underwent preoperative transradial coronary angiography (RA-CA). RESULTS: Baseline RA median diameters were 0.25 ± 0.04 cm in the cannulated arm and 0.22 ± 0.04 cm in the non-cannulated arm (P= 0.01). The flow-mediated dilatation (FMD) in the RA in the catheterized arm and in the control arm were 11.6 ± 7.9 and 14.2 ± 8.9 (P= 0.01), respectively. A statistically significant correlation was found between FMD of the catheterized RA and the time from RA-CA (Pearson's r= 0.348). Linear regression analysis confirmed that the FMD of the catheterized RA was dependent on days elapsed from the procedure (P= 0.032; OR 1.11, CI 0.009-0.203). Immunohistochemical evaluation showed extensive endothelial lesion in all examined RAs, with a trend towards reduction of the damage with time. Endothelial function and integrity of the cannulated arm did not reach those of the control arm in any of the study patients. CONCLUSIONS: RA-CA produces extensive damage to the RA. The lesions tend to heal with time but incomplete recovery of endothelial integrity and function is still present more than 30 days after the procedure. After RA-CA, the cannulated RA should not be used for CABG.