LANDSAT-4 sensor performance

1983 
Preflight and in-orbit sensor and data measurements indicate that TM meets or exceeds most specifications. Measured spectral band edges meet instrument specifications in 12 out of 14 cases; there is ample dynamic range. The signal-to-noise ratio exceeds specifications, except for band 3, channel 4; and band 7 channel 7 is very noisy but still meets specifications. The modulation transfer function of channel 4, band 2, is smaller than specified. Registration errors between the primary focal plane (PFP) and the cold focal plane (CFP) are about 0.75 pixels along-scan and 0.2 pixels across scan. Forward and reverse scan discontinuities, are well within ground-processing capabilities to rectify. Instrument gain variability, up to 7% for band 5, requires use of the internal calibration (IC) system to assure radiometric accuracy. Preliminary applications evaluation of image contents indicates that TM provides much better definition of edges than MSS.