Ultimate limits for the radiation hardness of silicon strip detectors for sLHC

2007 
The new SuperLHC upgrade will impose severe restrictions on the radiation hardness of silicon detectors since a maximum fluence of 10 16 particles/cm 2 is foreseen in the innermost region. Microstrip detectors have been fabricated in p-type high resistivity float zone silicon at CNM facilities, been irradiated at the TRIGA reactor in Ljubljana to a fluence of 10 16 neutrons/cm 2 and characterized at IFIC laboratory. The total collected charge before and after irradiation in the detectors has been measured by 90 Sr beta source and by infrared laser illumination. The results show that even after this extreme radiation fluence, p-type substrate detectors collect 3500 electrons when biased at 800 V, which is enough charge to induce a measurable signal with standard readout electronics. P-type strip detectors could be suitable for the middle and even inner regions of sLHC.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    7
    References
    3
    Citations
    NaN