Generalized min-max bound-based MRI pulse sequence design framework for wide-range T1 relaxometry: A case study on the tissue specific imaging sequence
2017
This paper proposes a new design strategy for optimizing MRI pulse sequences for T1 relaxometry. The design strategy optimizes the pulse sequence parameters to minimize the maximum variance of unbiased T1 estimates over a range of T1 values using the Cramer-Rao bound. In contrast to prior sequences optimized for a single nominal T1 value, the optimized sequence using our bound-based strategy achieves improved precision and accuracy for a broad range of T1 estimates within a clinically feasible scan time. The optimization combines the downhill simplex method with a simulated annealing process. To show the effectiveness of the proposed strategy, we optimize the tissue specific imaging (TSI) sequence. Preliminary Monte Carlo simulations demonstrate that the optimized TSI sequence yields improved precision and accuracy over the popular driven-equilibrium single-pulse observation of T1 (DESPOT1) approach for normal brain tissues (estimated T1 700–2000 ms at 3.0T). The relative mean estimation error (MSE) for T1 estimation is less than 1.7% using the optimized TSI sequence, as opposed to less than 7.0% using DESPOT1 for normal brain tissues. The optimized TSI sequence achieves good stability by keeping the MSE under 7.0% over larger T1 values corresponding to different lesion tissues and the cerebrospinal fluid (up to 5000 ms). The T1 estimation accuracy using the new pulse sequence also shows improvement, which is more pronounced in low SNR scenarios.