On-wafer Magnetic Resonance of Magnetite Nanoparticles

2015 
Abstract Magnetic resonance measurements of ferumoxytol and TEMPO were made using an on-wafer transmission line technique with a vector network analyzer, allowing for broadband measurements of small sample volumes (4 nL) and small numbers of spins (1 nmol). On-wafer resonance measurements were compared with standard single-frequency cavity-based electron paramagnetic resonance (EPR) measurements using a new power conservation approach and the results show similar line shape. On-wafer magnetic resonance measurements using integrated microfluidics and microwave technology can significantly reduce the cost and sample volumes required for EPR spectral analysis and allow for integration of EPR with existing lab-on-a-chip processing and characterization techniques for point-of-care medical diagnostic applications.