Development of deep-UV sensitive MPPC for liquid xenon scintillation detector

2015 
Abstract The liquid xenon (LXe) γ-ray detector for the MEG II experiment is based on a highly granular scintillation readout with Multi-Pixel Photon Counters (MPPCs). Here we report on the development of a large-area MPPC sensitive to LXe scintillation light in deep-UV range for the MEG II LXe detector. A prototype model of the deep-UV MPPC with an active area of 12 × 12 mm 2 was successfully tested in LXe, showing an excellent performance such as a high photon detection efficiency for the LXe scintillation light (about 20% at Δ V = 2.5 V ), a high internal gain (about 10 6 at Δ V = 2.5 V ) and an excellent single photoelectron resolution. The sensor chip of the MPPC is segmented into four sectors, which are then connected in series in order to reduce the overall sensor capacitance. The signal fall time of about 135 ns for the non-segmented sensor was significantly reduced down to 25 ns with the series-connected sensor segments. The preliminary results on the performance of the deep-UV MPPC are presented.