A New Calibration Transformer and Measurement Setup for Bridge Standard Calibrations Up To 5 kHz

2017 
This paper presents an improved calibration setup for static bridge standards with a carrier frequency supply voltage in the frequency range between 225 Hz and 5 kHz. In the past, the small bridge differential voltage in the mV/V range was calibrated in two subsequent voltage ratio calibrations. However, this procedure is time-consuming and only works reliably if both voltages are stable over time. For this purpose, a new setup was developed that is able to measure the differential voltage of the bridge standard in one single measurement. This reduces the calibration time by approximately two thirds, thereby also reducing the uncertainty.