LC-MS/MS method validation for determination of selected neonicotinoids in groundwater for the purpose of a column experiment
2019
The work was carried out to develop and validate a method based on liquid chromatography tandem mass spectrometry (LC-MS/MS) for the simultaneous detection and quantification of five neonicotinoid insecticides: acetamiprid, clothianidin, imidacloprid, thiacloprid and thiamethoxam in groundwater samples for the purpose of a further column experiment. This experiment will be used to analyze breakthrough curves of neonicotinoids under a saturated water flow to set transport parameters, where the reliability of the results of chemical analysis plays an important role. The validation was performed in a concentration range from 0.09 µg L⁻¹ to 100 µg L⁻¹ using an Infinity 1290 (Agilent, USA) chromatograph coupled with a QTrap 5500 mass spectrometer (Sciex, Canada). The calibration curves were obtained on the basis of the results of six standard solution analyses. Linearity was not lower than 0.998. The limit of detection was set at the lowest concentration which can be determined with an acceptable accuracy and precision. All pesticides had recoveries in the range 85–109% with relative standard deviation values less than 8.1%. The estimated measurement uncertainty did not exceed 30%, so the LC-MS/MS method fits for the intended purpose.