Distortion improvement of capsule endoscope image
2007
Distortion exists in the present capsule endoscope image resulting from the confined space and the wide-angle
requirement [8]. Based on the previous two lens works, the optimal design had obtained that the field of view was about
86 degrees , and MTF was about 18% at 100 lp/mm, but distortion would go to -26%. It's difficult to add another lens on
the 7mm optical path between the dome and imaging lenses for improving distortion. In order to overcome this problem,
we intend to design the optical dome as another optical lens. The original dome is transparent and has an equal thickness,
namely without refracting light almost. Our objective in this paper is to design the inner curvature of the dome and
associate two aspheric imaging lenses in front of the CMOS sensors to advance the distortion with maintaining field of
view and MTF under the same capsule volume. Furthermore, the paper proposes the real object plane of intestine is
nearly a curved surface rather than an ideal flat surface. Taking those reasons under consideration, we design three
imaging lenses with curved object plane and obtain the field of view is about 86 degrees , MTF is about 26% at 100
lp/mm, and the distortion improve to -7.5%. Adding the dome lens is not only to enhance the image quality, but also to
maintain the tiny volume requirement.