Automated Emergency Ventilation Devices in a Simulated Unprotected Airway

2009 
Abstract Background: Automated ventilation devices are becoming more popular for emergency ventilation, but there is still not much experience concerning the optimal ventilation mode. Methods: In a bench model representing a non-intubated patient in respiratory and cardiac arrest, we compared a pressure-cycled with a time- and volume-cycled automated ventilation device in their completely automated modes. The main study endpoints were inspiratory time, respiratory rate, stomach inflation, and lung tidal volumes. Results: The pressure-cycled device inspired for 6.7 s in the respiratory arrest setting (respiratory rate 5.6/min), and never reached its closing pressure in the cardiac arrest setting (respiratory rate 1 breath/min). The time- and volume-cycled device inspired in both settings for 1.7 s (respiratory rate 13 breaths/min). In the respiratory arrest setting, mask leakage was 620 ± 20 mL for the pressure-cycled device vs. 290 ± 10 mL for the time- and volume-cycled device ( p p p p p p