2-D/3-D image registration of implanted knee DR images with Kalman filter

2009 
Total knee arthroplasty (TKA) is an orthopedic surgery which replaces the damaged knee joint with the artificial one. To diagnose the function of the implanted knee joint, it is effective to estimate 3-D knee kinematics in vivo. There are some conventional methods for estimating kinematics of the implanted knee using 2-D/3-D image registration for X-ray fluoroscopic images and 3-D geometrical models of the knee implant. However, these methods are based on static image analysis although the knee joint continuously moves. This paper proposes an analysis method of the knee kinematics using digital radiography images with Kalman filter. Use of Kalman filter enables us to take into account the continuous knee movement. The experimental results showed that the proposed method estimated the knee joint angles within a mean error of 0.31 deg.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    8
    References
    1
    Citations
    NaN
    KQI
    []