CERAMIC-CERAMIC: ARTICULATIONS OF LONGEVITY FOR THE YOUNG ACTIVE PATIENT – AFFIRMS

2018 
Ceramic-on-ceramic bearings provide a solution to the osteolysis seen with traditional metal-on-polyethylene bearings. Sporadic reports of ceramic breakage and squeaking concern some surgeons and this bearing combination can show in vivo signs of edge loading wear which was not predicted from in vitro studies. Taper damage or debris in the taper between the ceramic and metal may lead to breakage of either a ceramic head or insert. Fastidious surgical technique may help to minimise the risk of ceramic breakage. Squeaking is usually a benign complication, most frequently occurring when the hip is fully flexed. Rarely, it can occur with each step of walking when it can be sufficiently troublesome to require revision surgery. The etiology of squeaking is multifactorial in origin. Taller, heavier and younger patients with higher activity levels are more prone to hips that squeak. Cup version and inclination are also relevant factors. Osteolysis following metal-on-UHMW polyethylene Total Hip Arthroplasty (THA) ...