Noninvasively Imaging pH at the Surface of ImplantedOrthopedic Devices with X‑ray Excited Luminescence ChemicalImaging

2019 
Implanted medical device-associated infections are a leading cause of fixation failure, and early diagnosis is the key to successful treatment. During infection, acidosis near the implant plays a role in antibiotic resistance and low pH is a potential infection indicator. Herein, we describe a pH sensor which attaches to the implants to noninvasively image local pH with high spatial resolution. The sensor has two layers: a scintillator layer which emits 620 and 700 nm light upon X-ray irradiation and a pH indicator layer containing bromocresol green dye that absorbs 620 nm luminescence in neutral/basic pH and passes 700 nm light at all pHs. We also developed a dedicated imaging system capable of scanning relatively large specimens through thick tissues. A focused X-ray beam irradiates one spot on the sensor, and the 620 to 700 nm peak ratio is measured to determine the local pH; images are acquired by scanning the X-ray beam across the surface and measuring the pH point-by-point. The sensor was covered wi...