Measurement of systolic and diastolic flow rates in the coronary artery system by x-ray densitometry.

1983 
The reliability of a modified videodensitometric and photodensitometric sampling technique for measuring phasic flow rates in the coronary artery system was examined. Electromagnetic flow measurements were performed in a circulatory model with continuous and pulsatile flow and intraoperatively in aortocoronary bypass grafts; cineangiograms were made simultaneously. Based on the front velocities of injected boluses of contrast medium, the densitometric measurement overestimated the electromagnetically measured flow systematically by about 20%. Systolic and diastolic flow rates in aortocoronary bypass grafts and coronary arteries determined from biplane cineangiograms in 34 patients generally revealed the typical pulsatile flow pattern familiar from electromagnetic and ultrasonic flow measurements. Flow velocities in unstenosed coronary arteries were nearly identical before and after branchings of the vessels, whereas the corresponding flow rates were higher in proximal than in distal segments. The identica...