An implantable radiation dosimeter for use in external beam radiation therapy
2004
An implantable radiationdosimeter for use with external beam therapy has been developed and tested both in vitro and in canines. The device uses a MOSFETdosimeter and is polled telemetrically every day during the course of therapy. The device is designed for permanent implantation and also acts as a radiographic fiducial marker. Ten dogs (companion animals) that presented with spontaneous, malignant tumors were enrolled in the study and received an implant in the tumor CTV. Three dogs received an additional implant in collateral normal tissue.Radiation therapy plans were created for the animals and they were treated with roughly 300 cGy daily fractions until completion of the prescribed cumulative dose. The primary endpoints of the study were to record any adverse events due to sensor placement and to monitor any movement away from the point of placement. No adverse events were recorded. Unacceptable device migration was experienced in two subjects and a retention mechanism was developed to prevent movement in the future. Daily dose readings were successfully acquired in all subjects. A rigorous in vitro calibration methodology has been developed to ensure that the implanted devices maintain an accuracy of ±3.5% relative to an ionization chamber standard. The authors believe that an implantable radiationdosimeter is a practical and powerful tool that fosters individualized patient QA on a daily basis.