Advancement of atmospheric-vacuum interfaces for mass spectrometers with a focus on increasing gas throughput for improving sensitivity.
2015
Ion sampling from an electrospray ionization (ESI) source was improved by increasing gas conductance of the MS inlet by 4.3-fold. Converting the gas throughput (Q) into sensitivity improvement was dependent on ion desolvation and handling of the gas load. Desolvation was addressed by using a novel slot shaped inlet that exhibited desolvation properties identical to the 0.58 mm i.d capillary. An assay tailored for “small molecules” at high chromatographic flow rate (500 μL/min) yielded a compound dependent 6.5 to 14-fold signal gain while analysis at nano chromatographic flow rate (300 nL/min) showed 2 to 3.5-fold improvement for doubly charged peptides. Improvement exceeding the Q (4.3-fold) at high chromatographic flow rate was explained by superior sampling of the spatially dispersed ion spray when using the slot shaped capillary. Sensitivity improvement across a wide range of chromatographic flow rate confirmed no compromise in ion desolvation with the increase in Q. Another improvement included less o...