Compact Method to Calibrate Frequency Response of Ultrasonic Transducer

2010 
Frequency response character of ultrasonic transducer often influences on the test signal partly, and then test errors arise. To the problem, a compact method is proposed to calibrate the frequency response in this paper. The experiment data was obtained in water- immerging test of the transducers, and a discrete transfer function is established based on system identification algorithms and then used for transducer calibration. The method is validated effective by experiment. Not only can the characteristic of transducers be indicated, but also a referenced method is presented for calibrating frequency response of LTI system.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader