New approach to resolve the humidity problem in VOC determination in outdoor air samples using solid adsorbent tubes followed by TD-GC–MS
2017
Abstract This study describes the humidity effect in the sampling process by adsorbent tubes followed by thermal desorption and gas chromatography–mass spectrometry (TD-GC–MS) for the determination of volatile organic compounds (VOCs) in air samples and evaluates possible solutions to this problem. Two multi-sorbent bed tubes, Tenax TA/Carbograph 1TD and Carbotrap B/Carbopack X/Carboxen 569, were tested in order to evaluate their behaviour in the presence of environmental humidity. Humidity problems were demonstrated with carbon-based tubes, while Tenax-based tubes did not display any influence. Silica gel, a molecular sieve and CaCl 2 were tried out as materials for drying tube to remove air humidity, placed prior to the sampling tube to prevent water from entering. The pre-tubes filled with 0.5 g of CaCl 2 showed the best results with respect to their blanks, the analytes recoveries and their ability to remove ambient humidity. To avoid the possible agglomeration of CaCl 2 during the sampling process in high relative humidity atmospheres, 0.1 g of diatomaceous earth were mixed with the desiccant agent. The applicability of the CaCl 2 pre-tube as drying agent prior to Carbotrap B/Carbopack X/Carboxen 569 tubes was tested in urban and industrial locations with samplings of air at high relative humidity. In addition, the results were compared with those obtained using Tenax TA/Carbograph 1TD tubes.