Optimized ultra performance liquid chromatography tandem high resolution mass spectrometry method for the quantification of paraquat in plasma and urine.
2016
Abstract A simple, sensitive and specific ultra performance liquid chromatography coupled to electrospray tandem high resolution mass spectrometry (UPLC-ESI-HRMS/MS) method has been developed and validated for quantification of paraquat in plasma and urine. The sample preparation was carried out by one-step protein precipitation with acetonitrile. The paraquat was separated with a HILIC column in 10 min. Detection was performed using Q Exactive Orbitrap mass spectrometer by Targeted-MS/MS scan mode. Methodological parameters, such as ammonium formate concentration, formic acid concentration, spray voltage, capillary temperature, heater temperature and normalized collision energy were optimized to achieve the highest sensitivity. The calibration curve was linear over the concentration range of LOQ-1000 ng/mL. LOD was 0.1 and 0.3 ng/mL, LOQ was 0.3 and 0.8 ng/mL for urine and plasma, respectively. The intra- and inter-day precisions were
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
37
References
14
Citations
NaN
KQI