Impact of HIV Type 1 Subtype Variation on Viral RNA Quantitation

1999 
We evaluated the performance of three HIV-1 RNA quantitation methods (Amplicor HIV-1 MONITOR-1.0, NASBA, and Quantiplex HIV RNA 2.0 [branched DNA (bDNA)]) using plasma specimens (N = 60) from individuals from Asia and Africa infected with one of three HIV-1 subtypes (A, Thai B [B'] or E; N = 20 each). Our results demonstrate that of the 20 subtype A specimens, 19 were quantifiable by the bDNA assay compared with 15 by the MONITOR-1.0 and 13 by NASBA. Of those quantifiable, the mean log10 difference was 0.93 between bDNA and MONITOR-1.0 and 0.46 between bDNA and NASBA. For subtype B' specimens, the correlation among methods was better with only 2 specimens missed by NASBA and 3 by the bDNA assay. However the missed specimens had viral burden near the lower limit (1000 copies/ml) for these assays. For the 20 subtype E specimens, MONITOR-1.0 and NASBA quantified RNA in 17 and 14 specimens, respectively, as compared with 19 specimens quantified by the bDNA assay. The correlation among different assays, especi...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    59
    References
    116
    Citations
    NaN