Detection of heart rate and rhythm with a smartphone-based electrocardiograph versus a reference standard electrocardiograph in dogs and cats
2016
OBJECTIVE To evaluate the diagnostic utility of ECGs acquired with a smartphone-based device, compared with reference 6-lead ECGs, for identification of heart rate and rhythm in dogs and cats. DESIGN Prospective study. ANIMALS 51 client-owned dogs and 27 client-owned cats. PROCEDURES Patients examined by a small animal referral cardiology service between April 2012 and January 2013 were enrolled consecutively. In each patient, a 30-second ECG was simultaneously acquired with a smartphone-based device (a bipolar, single-lead recorder coupled to a smartphone with an ECG application) and a standard 6-lead ECG machine. Recordings were evaluated by 3 board-certified cardiologists, and intra- and interobserver agreement were evaluated for both rhythm diagnosis and QRS polarity identification. RESULTS Values for instantaneous and mean heart rates for the smartphone-acquired and reference ECGs were within 1 beat of each other when mean heart rates were calculated. Intraobserver agreement for rhythm assessment was...