Comparison of in vitro hydrolysis, subcutaneous and intramedullary implantation to evaluate the strength retention of absorbable osteosynthesis implants

1990 
Abstract To evaluate the behaviour of mechanical properties of absorbable osteosynthesis implants in vivo , the strength retention of self-reinforced polyglycolide rods in distilled water at 37°C in the subcutis and femoral medullary cavity of the rabbit was investigated. The self-reinforced polyglycolide rods lost their strength significantly faster in vivo than In vitro . The strength loss of the self-reinforced polyglycolide rods was only slightly faster in the medullary cavity than in the subcutis. As the removal of the implants from the medullary cavity became difficult 5–6 wk after implantation, it is suggested that subcutaneous implantation would be a suitable method to evaluate the strength retention of absorbable osteosynthesis implants.