Gaps in the Traceability Chain of Human Growth Hormone Measurements
2013
BACKGROUND: Human growth hormone (hGH) is measured for the diagnosis of secretion disorders. These measurements fall under the EU Directive 98/79/EC on in vitro diagnostic medical devices requiring traceability of commercial calibrator values to higher-order reference materials or procedures ( Off J Eur Communities 1998 Dec 7;L 331:1–37). External quality assessment schemes show large discrepancies between results from different methods, even though most methods provide results traceable to the recommended International Standard (IS 98/574). The aim of this study was to investigate possible causes for these discrepancies.
METHODS: We investigated the commutability and recovery of hGH in reconstituted IS 98/574. We tested different reconstitution protocols and used 4 different serum matrices for spiking. These IS preparations were measured together with serum samples. We quantified hGH by 5 different methods in 4 different laboratories.
RESULTS: Results from the different methods correlated well for the serum samples. Mean discrepancies between results from different methods were ≤20%. None of the IS preparations was commutable for all the method comparisons. The recovery of hGH in preparations of IS 98/574 depended on the reconstitution protocol (>10-fold differences) and background matrix (relative differences ≤17% for different serum matrices).
CONCLUSIONS: The use of different protocols for reconstitution and spiking of hGH reference preparations affects quantification by immunoassays, potentially leading to a bias between commercial methods, despite the use of calibrators with values claimed to be traceable to the same higher-order reference material.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
23
References
16
Citations
NaN
KQI