123I-MIBG myocardial scintigraphy for the evaluation of Lewy body disease: are delayed images essential? Is visual assessment useful?

2016 
Objective:We performed 123I-meta-iodobenzylguanidine (123I-MIBG) myocardial scintigraphy for the diagnosis of Lewy body disease (LBD) and assessed whether the early heart-to-mediastinum (H/M) ratio was diagnostic and whether visual image analysis was useful.Methods:Our study included 453 patients with clinically suspected LBD who had undergone 123I-MIBG myocardial scintigraphy. We evaluated semi-quantitative values (early and delayed H/M ratio, washout rate) and performed visual image analysis. The sensitivity, specificity, accuracy, positive-predictive value (PPV) and negative-predictive value (NPV) of both analyses were calculated.Results:The early H/M ratio exhibited the best diagnostic ability. Its overall sensitivity, specificity, accuracy, PPV and NPV were 72.2%, 93.1%, 84.3%, 88.5% and 82.2%, respectively (cut-off value 2.2 or less). However, there was no significant difference from the delayed H/M ratio or the results of visual assessment.Conclusion:123I-MIBG myocardial scintigraphy accurately dif...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    23
    References
    9
    Citations
    NaN
    KQI
    []