A coupled-enzyme equilibrium method for measuring urea in serum: optimization and evaluation of the AACC study group on urea candidate reference method.
1980
We describe a coupled-enzyme equilibrium method for measuring urea in serum, which is performed on supernates prepared by treating each specimen with Ba(OH)2 and ZnSO4 (Somogyi reagent). Analytical recovery of [14C]urea added to a variety of matrices was essentially complete (mean, 100.6%) for the supernates after precipitation. Nine variables were univariately examined in arriving at the reaction conditions for the method: glutamate dehydrogenase, urease, 2-oxoglutarate, ADP, Tris . HCI, NADH, EDTA, pH, and temperature. The reagent is stable for at least 48 days at--20 degrees C and for 23 days at 4 degrees C. Mean analytical recovery of urea (14 mmol/L) added to seven different specimens (three different matrices) was 100.8%. The analytical linear range of the method extends to 30 mmol of urea per liter. Of 22 potential interferents, only bilirubin at 1 mmol/L (580 mg/L), hemoglobin at 10 g/L, and hydroxyurea at 6 mmol/L showed more than 2% interference. We discuss precision and effects of specimen dilution, and compare results for 100 human serum specimens with those measured for the same specimens with four other urea methods. We examined the effects of measuring a blank, consisting of sample and reagent without urease, with each specimen.