Effect of lubricants on friction in laboratory tests of a total disc replacement device

2013 
Some designs of total disc replacement devices have articulating bearing surfaces, and these devices are tested in vitro with a lubricant of diluted calf serum. It is believed that the lubricant found in total disc replacement devices in vivo is interstitial fluid that may have properties between that in Ringer’s solution and diluted calf serum. To investigate the effect of lubricants, a set of friction tests were performed on a generic model of a metal against metal ball-and-socket total disc replacement device. Two devices were tested: one with a ball radius of 10 mm and other with a ball radius of 16 mm; each device had a radial clearance of 0.015 mm. A spine simulator was used to measure frictional torque for each device in axial rotation, flexion–extension and lateral bending at frequencies of 0.25–2 Hz, under 1200 N axial load. Each device was tested with two different lubricants: a solution of new born calf serum diluted with deionised water and Ringer’s solution. The results showed that the fricti...