Contrast-Enhanced MR Angiography at 1.5 T After Implantation of Platinum Stents: In Vitro and In Vivo Comparison with Conventional Stent Designs
2005
OBJECTIVE. The objective of our study was to evaluate the in vitro and in vivo 3D contrast-enhanced MR angiography characteristics of a new platinum-based balloon-expandable stent system and compare this system with a variety of competing metallic stents.MATERIALS AND METHODS. All experiments were performed on 1.5-T scanners. In vitro experiments were performed using 10 stents implanted into a custom-built phantom. Different orientations of the stents along the magnetic field and multiple flip angles were examined. In addition, 19 patients underwent contrast-enhanced MR angiography after the implantation of 36 stents, including four patients with six platinum stents. Angiographic correlation was available for all 19 patients, and luminal patency and stent-induced artifacts were assessed quantitatively.RESULTS. Of the tested balloon-expandable stents, only the platinum-based stents created artifact causing luminal narrowing of 30% or less. All other balloon-expandable stents induced larger artifacts that r...