Differential Mobility Spectrometry Coupled with Multiple Ion Monitoring in Regulated LC-MS/MS Bioanalysis of a Therapeutic Cyclic Peptide in Human Plasma

2016 
A differential mobility spectrometry (DMS) in combination with a multiple ion monitoring (MIM) method was developed and validated for quantitative LC-MS/MS bioanalysis of pasireotide (SOM230) in human plasma. Pasireotide, a therapeutic cyclic peptide, exhibits poor collision-induced dissociation (CID) efficiency for multiple reaction monitoring (MRM) detection. Therefore, in an effort to increase the overall sensitivity of the assay, a DMS-MIM approach was explored. By selecting the most abundant doubly charged precursor ion in both the Q1 and Q3 of the mass analyzer in MIM and combining the DMS capability to significantly reduce the high matrix/chemical background noise, this new LC-DMS-MIM method overcomes the sensitivity challenge in the typical MRM method due to poor CID fragmentation of the analyte. Human plasma was spiked with pasireotide with concentrations in the range 0.01–50 ng/mL. Weak cation-exchange solid-phase extraction was employed for sample preparation. The sample extracts were analyzed ...