Primary standardisation of 204Tl using the efficiency tracing method.

2004 
Abstract The electron-capture and β − -emitting radionuclide 204 Tl has been known from previous experience to show discrepancies between different methods of standardisation. Source preparation is also difficult due to the complex chemistry of thallium. The Bureau International des Poids et Mesures (BIPM) intercomparison held in 1997, showed discrepancies of up to 10% between measurements of solid sources and liquid scintillation methods. These problems have been the subject of a BIPM CCRI(II) working group since 1999. This paper presents the results from a primary standardisation of a 204 Tl solution using 4π β − γ coincidence counting and liquid scintillation counting. The tracer technique was used for the 4 πβ − γ coincidence counting, where 204 Tl was traced with 134 Cs and 60 Co. The extrapolation to 100% β -efficiency was performed by three different approaches: foiling, adding carrier and measuring sources of different initial masses. The results showed that tracing with 60 Co and using external foiling gave the same result as obtained by liquid scintillation counting. A comparison of the results achieved by the different methods of measurement, tracers and methods for extrapolation is presented and discussed in this paper.