Amperometric Determination of Sulfide by Glassy Carbon Electrode Modified with Hemin Functionalized Reduced Graphene Oxide

2016 
This work describes a new approach for rapid detection of sulfide using a glassy carbon electrode (GCE) modified with hemin functionalized reduced graphene oxide (hemin-rGO). The modified GCE was applied to the selective amperometric detection of sulfide at a working potential of 0.05 V (vs. Ag/AgCl) at pH 7.5, which gave a linear range of 0.002–0.212 mM, a sensitivity of 246.5 µA mM−1 cm−2, and a detection limit of 1.3 µM. The proposed method was applied in determination of sulfide-spiked water, and the method was verified with recovery percentage of standard samples with an acceptable satisfaction.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    27
    References
    11
    Citations
    NaN