Influence of the laser repetition rate on the limits of detection in the femtosecond LIBS of the water solutions

2019 
The limit of detection for Mg, Mn, Sr, B was experimentally investigated depending on the laser pulse repetition rate in the time resolved femtosecond LIBS. Optical breakdown on the surface of aqueous solutions was generated by the femtosecond Ti:Sa laser with the following parameters: central wavelength - 800 nm, pulse width - 60 fs, pulse energy - 1 mJ, repetition rate 50 Hz, 166 Hz, 250 Hz, 500 Hz, 1000 Hz. The best limit of detection was obtained for the 166 Hz laser repetition rate and for the following elements Mg I (285.2 nm), Mn I (403 nm, 403.3 nm), B I (283.2 nm), Sr I (460.7 nm) are 0.058, 0.025, 1.13, 4.69-3 mg/kg respectively.