A Comparison of Point-of-Care Instruments Designed for Monitoring Oral Anticoagulation with Standard Laboratory Methods
2000
Our study compared point-of-care (POC) device monitoring with traditional clinical laboratory methods device of patients on oral anticoagulant therapy. The POC devices used in the study were Coumatrak, CoaguChek, CoaguChek Plus, Thrombolytic Assessment System (TAS) PT-One, TAS PTNC, TAS PT, Hemachron Jr. Signature, Protime Microcoagulation System, and Medtronics ACT II. The clinical laboratory method used thromboplastins with different ISI values: Innovin and Thromboplastin C Plus (TPC). All POC INRs showed strong correlation with both laboratory methods, with correlation coefficients of >0.900. All POC methods demonstrated a significant (p