Comparisons of the GlideScope and Macintosh Laryngoscope in Tracheal Intubation by Medical Students on Fresh Human Cadavers

2009 
Abstract Objective The GlideScope Video Laryngoscope (GS) is an intubating device that provides equal or better glottic views than conventional laryngoscopes, but correct tube placement is more time-consuming, even when performed by experienced operators. The aim of this study was to investigate the use of the GS compared with the more conventional Macintosh laryngoscope in easy and difficult tracheal intubation when performed by inexperienced medical students on fresh human cadavers Patients and Methods Forty-one medical students were assigned to perform tracheal intubation using the direct Macintosh laryngoscope (DL) and the GS. Each student was given four attempts, with a maximum of 180 seconds for each attempt, to successfully intubate the trachea with a 6.5-mm tracheal tube in each of two scenarios, one with an easy airway and the other with a difficult airway cadaver. Results The total time of intubation for the easy airway cadaver was significantly longer in the GS group (61.4 ± 4.8 seconds vs. 40.6 ± 5.3 seconds; p vs. 98.7 ± 10.2 seconds; p Conclusion Most inexperienced operators found the GS to be more time-consuming for tracheal intubation than DL in the easy airway cadaver. However, an obvious advantage was demonstrated when the GS was used for the difficult airway.