Radiation exposure in spine surgery using an image-guided system based on intraoperative cone-beam computed tomography: analysis of 107 consecutive cases.

2016 
OBJECTIVE The O-arm system in spine surgery allows greater accuracy, lower rate of screw misplacement, and reduced surgical time. Some concerns have been postulated regarding the radiation doses to patients and surgeons. To the best of the authors' knowledge, most of the studies in the literature were performed with the use of phantoms. The authors present data regarding radiation exposure of the surgeon and operating room (OR) staff in a consecutive series of patients undergoing spine surgery. METHODS Radiation exposure data were collected in a series of 107 patients who underwent spine surgery using the O-arm system. The doses received by the surgeon and the staff were collected using electronic dosimeters. RESULTS All patients underwent 1–3 scans. The mean radiation dose to the patients was 5.15 mSv (range 1.48–7.64 mSv). The mean dose registered for the scan operator was 0.005 μSv (range 0.00–0.03 μSv) while the other members of the surgical team positioned outside the OR received 0 μSv. CONCLUSIONS T...