Quality assessment of pharmaceutical tablet samples using Fourier transform near infrared spectroscopy and multivariate analysis

2017 
Abstract Determination of the content uniformity, assessed by the amount of an active pharmaceutical ingredient (API), and hardness of pharmaceutical materials is important for achieving a high-quality formulation and to ensure the intended therapeutic effects of the end-product. In this work, Fourier transform near infrared (FT-NIR) spectroscopy was used to determine the content uniformity and hardness of a pharmaceutical mini-tablet and standard tablet samples. Tablet samples were scanned using an FT-NIR instrument and tablet spectra were collected at wavelengths of 1000–2500 nm. Furthermore, multivariate analysis was applied to extract the relationship between the FT-NIR spectra and the measured parameters. The results of FT-NIR spectroscopy for API and hardness prediction were as precise as the reference high-performance liquid chromatography and mechanical hardness tests. For the prediction of mini-tablet API content, the highest coefficient of determination for the prediction ( R 2 p ) was found to be 0.99 with a standard error of prediction (SEP) of 0.72 mg. Moreover, the standard tablet hardness measurement had a R 2 p value of 0.91 with an SEP of 0.25 kg. These results suggest that FT-NIR spectroscopy is an alternative and accurate nondestructive measurement tool for the detection of the chemical and physical properties of pharmaceutical samples.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    28
    References
    10
    Citations
    NaN
    KQI
    []