Evaluation of the load system produced by a single intrusion bend in a maxillary lateral incisor bracket with different alloys

2018 
ABSTRACT Objectives: To evaluate if a 0.5-mm vertical bend applied on an incisor bracket produces movements in other planes and if different wires influence these effects. Materials and Methods: An acrylic model of a treated patient with brackets passively bonded was attached to an Orthodontic Force Tester, and a load cell was attached to the left lateral incisor. Thirty 0.019 × 0.025-inch archwires were divided into three groups according to their alloy: SS (stainless steel), B-Ti (beta-titanium), and MF (beta-titanium wire coated with nickel-titanium). Step-bends of 0.5 mm high were placed on the lateral incisor bracket using a universal plier, and the forces and moments in three dimensions were statistically analyzed by analysis of variance and Tukey post hoc test. Results: SS produced a larger force (3.4 N) than the B-Ti (1.41 N) and the MF (0.53 N; P < .001). Lingual forces were produced by the SS (0.82 N) and B-Ti (0.31 N) groups, while in the MF group, the force was insignificant. SS produced a mes...