ROC curve analysis of lesion detectability on phantoms : comparison of digital spot mammography with conventional spot mammography

2001 
Although conventional screen–film mammography has excellent spatial resolution and is commonly used as a screening tool, certain inherent limitations prevent its further improvement. New digital mammography techniques, despite lower spatial resolution than screen–film mammography, may overcome these limitations. This study compared lesion detectability between charge coupled device-based digital spot mammography and conventional spot mammography. A total of 100 sets of images of specially designed breast phantoms was acquired, with variable background achieved by overlapping several layers of grapefruit fibre on a 4 cm thick lucite slab, using both modalities. 75 sets were “normal” images and 25 sets were images with simulated lesions. Four radiologists assessed the images according to a five-point confidence scale. The results were used to construct receiver operating characteristic curves. No statistical difference was observed between the two sets of curves for individual radiologists as well as pooled...