Performance of High Stopping Power Bismuth-loaded Plastic Scintillators for Radiation Portal Monitors

2020 
Plastic scintillators are widely used in radiation portal monitors because of their low cost and availability in large sizes. However, due to their low density and low effective atomic number ( Z ), they offer low intrinsic efficiency and little spectroscopic information. The addition of high- Z constituents to these plastics can greatly increase both their total stopping power and the amount of photoelectric absorption, leading to full-energy deposition and thus useful gamma spectra. In this article, we present the performance of the latest formulation of Bi-loaded plastic scintillators showing their useful spectroscopic information up to relatively high energy (~1 MeV) due to their high stopping power compared to the current commercially available plastics. These Bi-loaded plastics use 20 weight percent (wt%) Bi-pivalate (8 wt% elemental Bi) dissolved in polyvinyltoluene (PVT) matrix and conventional fast fluors (~10 ns decay time). These Bi-loaded plastics achieve up to approximately 6000 photons/MeV and have been produced in sizes up to 17 in3. The performance of these Bi-loaded plastics is also demonstrated in the existing portal monitor hardware (Rapiscan Model TSA Trainer 770) showing the possibility to provide improved sensitivity as a drop-in replacement with continued scale-up.