Background—Fractional flow reserve (FFR), an index of the hemodynamic severity of coronary stenoses, is derived from invasive measurements and requires a pressure-monitoring guidewire and hyperemic stimulus. Angiography-derived FFR measurements (FFRangio) may have several advantages. The aim of this study is to assess the diagnostic performance and interobserver reproducibility of FFRangio in patients with stable coronary artery disease. Methods and Results—FFRangio is a computational method based on rapid flow analysis for the assessment of FFR. FFRangio uses the patient’s hemodynamic data and routine angiograms to generate a complete 3-dimensional coronary tree with color-coded FFR values at any epicardial location. Hyperemic flow ratio is derived from an automatic resistance-based lumped model of the entire coronary tree. A total of 203 lesions were analyzed in 184 patients from 4 centers. Values derived using FFRangio ranged from 0.5 to 0.97 (median 0.85) and correlated closely (Spearman ρ=0.90; P<0.0...