Comparison of lesion formation between contact force-guided and non-guided circumferential pulmonary vein isolation: A prospective, randomized study
2014
Background Contact force (CF) monitoring could be useful in accomplishing circumferential pulmonary vein (PV) isolation (CPVI) for atrial fibrillation (AF). Objective The purpose of this study was to compare procedure parameters and outcomes between CF-guided and non-guided CPVI. Methods Thirty-eight consecutive AF patients (mean age 60 ± 11 years, 28 paroxysmal AF) undergoing CPVI were randomized to non–CF-guided (n = 19) and CF-guided (n = 19) groups. CPVI was performed with the ThermoCool SmartTouch catheter in both groups. The end-point was bidirectional block between the left atrium (LA) and PV. In the CF group, CF was kept between 10 and 20 g during CPVI, whereas in the non-CF group, all CF information was blanked. Radiofrequency energy at 30 W in the anterior and 25 W in the posterior LA wall was applied for 20–25 seconds at each point. Results CPVI was successfully accomplished without any major complications in both groups. Mean CF in the non-CF and CF groups were 5.9 ± 4.5 g and 11.1 ± 4.3 g, respectively, for left-side CPVI, and 9.8 ± 6.6 g and 12.1 ± 4.8 g, respectively, for right-side CPVI (both P P P = NS), respectively. Total number of residual conduction gaps was 6.3 ± 3.0 in the non-CF group and 2.8 ± 1.9 in the CF group ( P P = .34). Conclusion CF-guided CPVI is effective in reducing procedure time and additional touch-up ablation and may improve long-term outcome.