ACCURACY OF CT-BASED CUSTOM-MADE SURGICAL TEMPLATING INSTRUMENTATION FOR PATELLAR RESECTION IN TOTAL KNEE ARTHROPLASTY: IN VITRO STUDY

2018 
Patella resection has been the least controlled element of total knee arthroplasty (TKA). We have developed an intraoperative guide system involving a custom-made surgical template designed on the basis of a three-dimensional computer simulation incorporating computed tomography (CT) data for several years. This time we have applied this intraoperative guide system for the patella resection in TKA. We investigated the accuracy of CT-based patient-specific templating (PST) for patella resection using cadaveric knee joints in vitro.To plan the corrective patella resection, we attempted to simulate a three-dimensional patella resection with the use of computer models of the patella. From CT images of the patella we obtained three-dimensional surface models of the patella by performing a three-dimensional surface generation of the bone cortex. After the patella resection using CT-based custom-made surgical templating instrumentation, CT scan was performed again and we compared the patella shape in three-dimen...