Endobronchial Navigation Guided by Cone-Beam CT–Based Augmented Fluoroscopy without a Bronchoscope: Feasibility Study in Phantom and Swine

2020 
Abstract Purpose To evaluate the accuracy of cone-beam computed tomography (CT)-based augmented fluoroscopy (AF) image guidance for endobronchial navigation to peripheral lung targets. Methods Prototypic endobronchial navigation AF software that superimposed segmented airways, targets, and pathways based on cone-beam CT onto fluoroscopy images was evaluated ex vivo in fixed swine lungs and in vivo in healthy swine (n = 4) without a bronchoscope. Ex vivo and in vivo (n = 3) phase 1 experiments used guide catheters and AF software version 1, whereas in vivo phase 2 (n = 1) experiments also used an endovascular steerable guiding sheath, upgraded AF software version 2, and lung-specific low-radiation-dose protocols. First-pass navigation success was defined as catheter delivery into a targeted airway segment solely using AF, with second-pass success defined as reaching the targeted segment by using updated AF image guidance based on confirmatory cone-beam CT. Secondary outcomes were navigation error, navigation time, radiation exposure, and preliminary safety. Results First-pass success was 100% (10/10) ex vivo and 19/24 (79%) and 11/15 (73%) for in vivo phases 1 and 2, respectively. Phase 2 second-pass success was 4/4 (100%). Navigation errors were 2.2 ± 1.2 mm ex vivo and 4.9 ± 3.2 mm and 4.0 ± 2.6 mm for in vivo phases 1 and 2, respectively. No major device-related complications were observed in the in vivo experiments. Conclusions Endobronchial navigation is feasible and accurate with cone-beam CT-based AF image guidance. AF can guide endobronchial navigation with endovascular catheters and steerable guiding sheaths to peripheral lung targets, potentially overcoming limitations associated with bronchoscopy.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    24
    References
    3
    Citations
    NaN
    KQI
    []