Assessing the accuracy of commercially available gas sensors for the measurement of ambient ozone and nitrogen dioxide

2018 
AbstractThe objective of the National Institute for Occupational Safety and Health (NIOSH) accuracy criterion is to ensure that measurements from monitoring devices are within ±25% of the true concentration of the analyte with 95% certainty. To determine whether NO2 and O3 sensors meet this criterion, three commercially available units (Cairclip O3/NO2, Aeroqual NO2, and Aeroqual O3 sensors) were co-located three times with validated instruments (NOx chemiluminescence [NO2mon] and photometric O3 analyzers [O3mon]) at an outdoor monitoring station. As cofactors of sensor performance such as temperature (T) and relative humidity (RH) potentially influence the response of NO2 and O3 sensors, corrections for cofactors were made by using T, RH, and the sensor measurements to predict measurements made by NO2mon and O3mon during the first co-location period (training dataset). The developed models were tested in the merged data obtained from the second and third co-location periods (testing dataset). In the trai...