AbstractAim. Mercury-in-silastic strain gauge pletysmography (SGP) is a well-established technique for blood flow and blood pressure measurements. The aim of this study was to examine (i) the possible influence of clinical clues, e.g. the presence of wounds and color changes during blood pressure measurements, and (ii) intra- and inter-observer variation of curve interpretation for segmental blood pressure measurements. Methods. A total of 204 patients with known or suspected peripheral arterial disease (PAD) were included in a diagnostic accuracy trial. Toe and ankle pressures were measured in both limbs, and primary observers analyzed a total of 804 pressure curve sets. The SGP curves were later reanalyzed separately by two observers blinded to clinical clues. Intra- and inter-observer agreement was quantified using Cohen's kappa and reliability was quantified using intra-class correlation coefficients, coefficients of variance, and Bland-Altman analysis. Results. There was an overall agreement regardin...